Teledyne Card Game M200EU User Manual

MANUAL ADDENDUM  
ULTRA SENSITIVITY MODEL  
M200EU  
NO/NO2/NOx ANALYZER  
(Addendum to M200E Manual PN 04410)  
Teledyne Advanced Pollution Instrumentation  
9480 Carroll Park Drive  
San Diego, CA 92121-5201  
USA  
Toll-free Phone: 800-324-5190  
Phone: 858-657-9800  
Fax: 858-657-9816  
05385 Rev C  
Copyright 2009-2010  
DCN 5731  
Teledyne Advanced Pollution Instrumentation  
14 May 2010  
 
 
Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer  
Addendum to M200E  
1. OVERVIEW  
The model M200EU is a close derivative of the Model 200E NOx Analyzer. The main  
differences are as follows.  
The M200EU is more sensitive at low NOx levels than a standard 200E. The  
instrument has a user selectable full scale range of 0-5 PPB to 0-2000 PPB of  
NOx.  
The reaction cell is gold plated. In addition, there is a pre-reactor, which allows  
the measurements of any hydrocarbon interferents in the sample stream.  
The sample flow is about 1000 cc/min.  
The M200EU is equipped with a high performance pump, capable of producing a  
reaction cell pressure of less than 5 Inch Hg absolute.  
In addition to the contents of this M200EU Addendum, you must use the M200E manual.  
CAUTION!  
The following procedures apply only to the External Zero Air Scrubber and NOT to the inline  
exhaust scrubber cartridge that is part of the pump pack assembly.  
NOTE  
The M200EU has equivalency approval, and may be used  
for EPA monitoring.  
TAPI 05385 Rev B (DCN5519)  
PRINTED DOCUMENTS ARE UNCONTROLLED  
1
 
 
Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer  
Addendum to M200E  
2. SPECIFICATIONS, AGENCY  
APPROVALS, AND WARRANTY  
Min/Max Range  
Min: 0-5 ppb of Full Scale  
(Physical Analog Output)  
Max 0-2000 ppb of Full Scale  
Measurement Units  
Noise at Zero1  
ppb, ug/m3 (user selectable Single, independent or auto ranges)  
<25 ppt RMS  
Noise at Span1  
Lower Detectable Limit1  
<0.25% RMS of reading above 50 ppb  
50 ppt RMS  
<0.1 ppb / 24 hours  
<0.2 ppb / 7 days  
Zero Drift2  
Span Drift2  
<0.5% of reading or 50 ppt RMS whichever is greater / 24 hrs  
<1% of reading or 100 ppt RMS whichever is greater / 7 days  
Linearity  
1% of full scale or 0.1 ppb whichever is greater  
0.5% of reading above 50 ppb  
20 sec  
Precision  
Lag Time1  
Rise / Fall Time1, 3  
95% in <50 sec  
Sample Flow Rate  
1000 cc/min. ± 10%  
20-30oC within drift and noise specifications, and operating with EPA equivalency  
5-35oC safe operating range  
Temperature Range  
Humidity  
0-95% RH non-condensing  
Propylene rejection ratio > 20,000:1  
Ethylene rejection ratio > 40,000:1  
Hydrocarbon Interference  
Voltage Coefficient  
Dimensions H x W x D  
Weight, Analyzer  
Weight, Ext Pump Pack  
Power, Analyzer  
Power, Analyzer4  
Power, Ext Pump  
Power, Ext Pump4  
Analog Outputs  
< 0.1% per V  
7" x 17" x 23.5" (178mm x 432 mm x 597 mm)  
18 kg (40 lbs)  
9.5 kg (21 lbs)  
100V~ 50/60 Hz, 120V~ 60 Hz, 230V~ 50 Hz, 2.5A, 125 watts  
230V~ 50 Hz, 2.5A  
110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts  
230V~ 50 Hz, 2.2A  
Four (4) Outputs  
All outputs: 100 mV, 1, 5, 10V  
Analog Output Ranges  
Three outputs convertible to 4-20 mA isolated current loop  
All ranges with 5% under/over range  
Analog Output Resolution  
Status Outputs  
1 part in 4096 of selected full-scale voltage  
8 Status Outputs from opto-isolators, 7 defined, 1 spare  
6 Control Inputs, 4 defined, 2 spare  
Control Inputs  
2
TAPI 05385 Rev B  
PRINTED DOCUMENTS ARE UNCONTROLLED  
 
 
Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer  
Addendum to M200E  
COM 1: RS-232;  
Serial I/O  
COM 2: RS-232 or RS-485 (Ethernet interface available)  
Baud Rate: 300-115200  
Environmental4  
Certifications  
Installation Category (Over-voltage Category) II, Pollution Degree 2  
US EPA: Reference Method Number RFNA 1194-099  
CE: EN61326 (1997 w/A1:98) Class A, FCC Part 15 Subpart B Section 15.107 Class  
A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 &A2;97) Class A  
(1)  
As defined by the USEPA  
(2)  
At constant temperature and voltage  
With adaptive filter, > 20 ppb change  
Electrical rating for CE mark compliance  
(3)  
(4)  
TAPI 05385 Rev B (DCN5519)  
PRINTED DOCUMENTS ARE UNCONTROLLED  
3
 
Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer  
Addendum to M200E  
3. PNEUMATICS  
Refer to the following pneumatic diagram. It shows the internal pneumatics of the  
M200EU in its standard configuration, along with the optional Zero/Span valves. The  
sample flow is about 1000 cc/min. Also during Auto Zero mode, the sample gas is routed  
through a pre-reactor before the mixture goes through the reaction cell. Since the reaction  
rate constant between NO and ozone is several time faster than the reaction rate constant  
between hydrocarbons and ozone, this mechanism helps to eliminate the interference  
effects of hydrocarbons in NOx measurements. In addition, the M200EU is equipped  
with a high performance pump, capable of producing a reaction cell pressure of less than  
5 Inch Hg absolute.  
Figure 3-1: Pneumatic Diagram With Zero/Span Option  
4
TAPI 05385 Rev B  
PRINTED DOCUMENTS ARE UNCONTROLLED  
 
 
Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer  
Addendum to M200E  
4. SPARE PARTS LIST  
Note: Only parts different from M200E are listed below. Refer to the M200E users  
manual for the common spare parts.  
Part Number  
011340800  
024600100  
011420700  
011440200  
041800600  
001270001  
001330001  
001350001  
014080200  
022880000  
Description  
Assy, Sensor, M200EU  
Assy, Prereactor NO/NOx valve, M200EU  
Assy, NOx cell, Gold plated, M200EU  
Assy, PMT, M200AU/EU  
PCA, PMT Preamp, VR, M200EU  
Manifold, Gold Plated, NOx  
Sleeve, Gold Plated, NOx  
Nozzle, Outer, Gold Plated  
Assy, HVPS, M200AU/EU  
Assy, HVPS, Low Leak Current, M100EU/M200EU  
TAPI 05385 Rev B (DCN5519)  
PRINTED DOCUMENTS ARE UNCONTROLLED  
5
 
 

Taylor Electric Grill AG 1360 BL User Manual
Toshiba Network Card AR5004G User Manual
Toshiba Scale EM1 31059 User Manual
Tripp Lite Surge Protector D10B2 User Manual
Tripp Lite TV Cables EZB 100 User Manual
Ultra Products Ventilation Hood ULT30240 User Manual
Uniden CD Player PC78 User Manual
United States Pumice Company Nebulizer 98580 User Manual
Universal Remote Control Universal Remote OCE 0009D User Manual
Vizio CRT Television VOJ370 User Manual